Skip to main content
Matrix One>Blog>21 CFR Part 11: What It Requires for Electronic Records and Signatures

21 CFR Part 11: What It Requires for Electronic Records and Signatures

Last updated on

21 CFR Part 11 is the FDA regulation that decides when electronic records and electronic signatures count as equal to paper records and ink signatures. Short answer: it asks for 11 controls on closed systems in section 11.10(a) to (k), three items shown with every signature under 11.50, a signature to record link that cannot be broken by ordinary means under 11.70, unique and identity checked signatures under 11.100, two part signing under 11.200 and five password controls under 11.300. Since FDA's 2003 scope guidance, the agency applies enforcement discretion to the validation, audit trail, retention and copying requirements, but Part 11 remains in effect and the records themselves must still be kept. Matrix Req is the Matrix One tool that signs documents under Part 11 with a full audit trail.

A note on our interest: we work at Matrix One, which builds Matrix Req, Matrix Quality and Matrix Validate, and all three sign records under Part 11. This page walks through the regulation section by section from the eCFR text we read on 10 November 2026, then covers what FDA enforces today, what changed recently and what you need in place. For ranked lists of tools, see our guides to the best 21 CFR Part 11 compliant validation software and the best eQMS with Part 11 electronic signatures and audit trails.

What is 21 CFR Part 11?

The name is a citation. Read it from left to right and it tells you where the rule sits in US federal law:

  • "21" is Title 21 of the Code of Federal Regulations, the title that covers food and drugs. The CFR has 50 titles in total.

  • "CFR" is the Code of Federal Regulations, the codified rules of the United States federal government. Title 21 has three chapters: Chapter I for the FDA, Chapter II for the Drug Enforcement Administration and Chapter III for the Office of National Drug Control Policy.

  • "Part 11" sits in Chapter I, Subchapter A. Its scope is electronic records and electronic signatures, which includes electronic submissions to FDA.

Part 11 itself is short. It has three subparts and ten sections, and every requirement on this page traces back to one of them:

SubpartSectionsWhat it covers
A, General Provisions11.1, 11.2, 11.3Scope, implementation and definitions
B, Electronic Records11.10, 11.30, 11.50, 11.70Closed and open system controls, signature manifestations, signature to record linking
C, Electronic Signatures11.100, 11.200, 11.300General signature rules, signature components, ID code and password controls

Who has to comply with 21 CFR Part 11?

Part 11 applies to pharmaceutical, biotechnology and medical device companies, and to anyone else creating, modifying, maintaining, archiving, retrieving or transmitting electronic records that an FDA regulation requires. That reaches clinical trial sponsors and sites, contract manufacturers, distributors and any supplier keeping regulated records for them.

The trigger is in section 11.1(b): Part 11 covers records in electronic form kept under any records requirement in agency regulations, the so called predicate rules, and electronic records submitted to FDA under the Federal Food, Drug, and Cosmetic Act and the Public Health Service Act, even when no regulation names them. If no predicate rule requires the record and you never submit it, Part 11 does not apply to it.

What does Subpart A, the general provisions, say?

Section 11.1, scope: Part 11 sets the criteria under which FDA treats electronic records and electronic signatures as trustworthy, reliable and generally equivalent to paper records and handwritten signatures. One exclusion surprises people: Part 11 does not apply to paper records that are, or have been, transmitted by electronic means, such as a signed paper form sent as an email attachment.

Section 11.1 also says computer systems, controls and their documentation must be readily available for FDA inspection. In practice that proof is the validation of the software that creates the records and signatures, plus the procedures around it.

Section 11.2, implementation: for records you must keep but do not submit, you may use electronic records and signatures in place of paper if Part 11 is met. For records you submit, the document type must also be one FDA lists in public docket No. 92S-0251 as accepted in electronic form. Section 11.3 holds the definitions, including closed system, open system, digital signature and biometrics.

What does Subpart B require for electronic records?

What controls does section 11.10 require on a closed system?

A closed system is one where access is controlled by the people responsible for the content of the electronic records on it. A validated SaaS tool your company administers is the usual example. Anyone using a closed system must run procedures and controls that protect the authenticity, integrity and, where appropriate, the confidentiality of the records, and that stop a signer from readily denying a signed record as not genuine. Section 11.10 lists eleven:

ClauseControlWhat it means in practice
11.10(a)ValidationProve the system is accurate, reliable, performs as intended and can spot invalid or altered records
11.10(b)CopiesProduce accurate, complete copies in human readable and electronic form for FDA review
11.10(c)Record protectionKeep records retrievable for the full retention period
11.10(d)System accessLimit access to authorised individuals
11.10(e)Audit trailsSecure, computer generated, time stamped trails of every create, modify and delete; changes never obscure earlier entries
11.10(f)Operational checksEnforce the permitted sequence of steps, for example review before approval
11.10(g)Authority checksOnly authorised people can use the system, sign, alter a record or perform the operation
11.10(h)Device checksCheck, where appropriate, that the source of data input is valid
11.10(i)Personnel qualificationPeople who build, maintain or use the system have the education, training and experience for it
11.10(j)AccountabilityWritten policies hold people responsible for actions under their signatures
11.10(k)Documentation controlControl distribution and use of system documentation, with change control on it

Note what the list mixes. Clauses (a) to (h) are mostly technical controls a software vendor can build. Clauses (i) to (k) are procedures only your organisation can write and follow, which is why no tool can make you Part 11 compliant on its own.

What extra controls does section 11.30 add for open systems?

An open system is one where access is not controlled by the people responsible for the records on it. Everything in 11.10 still applies, as appropriate, plus additional measures such as document encryption and appropriate digital signature standards to protect authenticity, integrity and confidentiality from the point a record is created to the point it is received.

What must every electronic signature show under section 11.50?

A signed electronic record must clearly show three things:

  • the printed name of the signer;

  • the date and time the signature was executed;

  • the meaning of the signature, such as review, approval, responsibility or authorship.

Section 11.50(b) adds that these three items fall under the same controls as the record and must appear in any human readable form of it, on screen or in a printout.

How must signatures be linked to records under section 11.70?

Electronic signatures, and handwritten signatures executed to electronic records, must be linked to their records so that they cannot be excised, copied or otherwise transferred to falsify an electronic record by ordinary means. A signature image pasted into a PDF fails this test, because anyone can copy it to another file.

What does Subpart C require for electronic signatures?

What are the general requirements in section 11.100?

  • Unique: each electronic signature belongs to one individual and is never reused by or reassigned to anyone else (11.100(a)).

  • Verified: the organisation verifies a person's identity before it establishes or sanctions their electronic signature (11.100(b)).

  • Certified to FDA: before or at the time of first use, you certify to FDA that the electronic signatures in your system are intended as the legally binding equivalent of handwritten signatures (11.100(c)).

The certification rule changed on 2 March 2023 (88 FR 13018). The letter must still be signed with a traditional handwritten signature, but it can now be submitted in electronic or paper form, and the regulation points to FDA's web page on Letters of Non-Repudiation Agreement for where to send it instead of the paper only route to a fixed FDA office address that applied before. Many older Part 11 guides still quote the pre 2023 wording.

How must non-biometric signatures be built under section 11.200?

  • They use at least two distinct identification components, such as an ID code and a password.

  • In one continuous period of controlled access, the first signing uses all components; later signings in the same session need at least one component that only the individual can execute.

  • Signings outside one continuous session each need all components.

  • They are used only by their genuine owners, and any attempt by someone else to use a signature requires two or more people to collaborate.

Biometric signatures must be designed so that nobody other than their genuine owner can use them (11.200(b)).

Which password controls does section 11.300 require?

  • No two individuals share the same combination of ID code and password (a).

  • ID code and password issuances are periodically checked, recalled or revised, for example through password ageing (b).

  • Lost, stolen or compromised tokens and cards are electronically deauthorised under loss management procedures, with controlled replacements (c).

  • Transaction safeguards prevent unauthorised use and report attempts urgently to the system security unit and, as appropriate, management (d).

  • Tokens and cards are tested at issue and periodically to confirm they work and have not been altered (e).

Is FDA still enforcing every part of 21 CFR Part 11?

Not to the letter. FDA's guidance Part 11, Electronic Records; Electronic Signatures, Scope and Application (September 2003, docket FDA-2003-D-0143, read on fda.gov on 10 November 2026) says FDA interprets the scope of Part 11 narrowly and does not intend to take enforcement action on the validation, audit trail, record retention and record copying requirements as explained in that guidance. It also applies enforcement discretion to legacy systems that were operating before 20 August 1997, the date Part 11 took effect.

Two limits matter. Part 11 remains in effect, and the records must still be kept and controlled under the predicate rules. And FDA keeps adding sector guidance on top: the final guidance Electronic Systems, Electronic Records, and Electronic Signatures in Clinical Investigations: Questions and Answers (October 2024, docket FDA-2017-D-1105) is written for sponsors, clinical investigators, institutional review boards and contract research organisations.

How does the QMSR change 21 CFR Part 11 for medical device makers?

Part 11 itself has not changed since the 2023 amendment to section 11.100; the eCFR version history shows no later edit. What changed is the predicate rule behind most device records. The Quality Management System Regulation took effect on 2 February 2026, amending 21 CFR Part 820 to incorporate ISO 13485:2016 by reference (FDA QMSR page, read 10 November 2026).

Part 820 is still an agency records requirement, so under section 11.1(b) every electronic QMS record it requires, from design and development files to complaint records, still falls under Part 11. The validation duty also now reads through ISO 13485 clause 4.1.6 for QMS software, alongside 11.10(a). Our guide on how to validate QMS software efficiently covers the method.

Which Matrix One tools sign records under 21 CFR Part 11?

Matrix Req lets a team route a document for review, sign it under 21 CFR Part 11 and keep a full audit trail, and it generates trace tables and design documents from live project data. Electronic signatures, ISO 27001:2022 certified cloud hosting and full history come in every plan, and we provide a validation package for Matrix Req itself with plans, reports, test scripts and traceability matrices (Matrix Req product page, read 10 November 2026). Teams use it for 510(k), PMA and CE mark submissions.

Matrix Quality, our eQMS, puts a full audit trail on every record with 21 CFR Part 11 and EU Annex 11 electronic signatures on every approval (Matrix Quality, read 10 November 2026). Matrix Validate signs validation plans, trace matrices and summary reports with Part 11 compliant e-signatures that require full credentials at the point of signing. As that page says, compliance also depends on your own procedures.

Our own data: Matrix Req customers complete the vendor validation pack in about 4 hours on average, a figure from our Matrix Req demos in April 2026. And real feedback from a US biotech, two years commercial, with about 20 GxP systems, asked for test runs recorded as they happen rather than signed on paper and scanned later, which is the 11.10(e) audit trail point in a buyer's words.

What Matrix Req is built for, and what you would buy alongside it

Matrix Req is built for medical device teams that write, trace and sign design and technical documentation in one place, with Jira, GitHub and GitLab issues linked from requirements. For process, cleaning and method validation run paperless on the shop floor, Kneat Gx is built for large life sciences companies and covers those validation types, which neither Matrix Req nor Matrix Validate offers (our Kneat comparison, facts checked 3 November 2026).

What do you need in place before Matrix Req signs Part 11 records?

  • A named owner for the requirements and document structure, and a decision on which items live in Jira and which in Matrix Req.

  • Your existing trace matrix exported, so links can be mapped on import, and the onboarding sessions booked.

  • An identity check for every user before their account can sign, as 11.100(b) requires.

  • Your 11.100(c) certification letter sent to FDA before or at first use of electronic signatures.

  • Written signature accountability policies and training records for users, covering 11.10(i) and 11.10(j).

  • The Matrix Req validation package run and signed off, so 11.10(a) has evidence behind it.

Summary: what does 21 CFR Part 11 require?

21 CFR Part 11 requires that any electronic record kept or submitted under an FDA rule is created in a validated, access controlled system with secure audit trails, and that every electronic signature is unique, identity checked, certified to FDA, shown with name, date and meaning, and permanently linked to its record. FDA still applies its 2003 enforcement discretion to validation, audit trail, retention and copying, but the rule and the predicate records remain in force, and the QMSR keeps device records inside it. Matrix Req is the Matrix One tool that signs those documents with a full audit trail. The three things that matter most:

  1. The 11 closed system controls in 11.10, including validation (a) and audit trails (e), part technical and part procedural.

  2. Signature manifestation and linking under 11.50 and 11.70: name, date and time, meaning, and no copy and paste signatures.

  3. Signature identity under 11.100 to 11.300: unique, verified, certified to FDA, two components and controlled passwords.

To compare tools against these rules, see our ranked guide to the best 21 CFR Part 11 compliant validation software.

Last updated: 10 November 2026.

Watch: 21 CFR Part 11 compliant software, what your eQMS must do

What Part 11 actually asks of electronic records and signatures, and how Matrix Quality covers it out of the box.

Questions buyers ask about 21 CFR Part 11

Is your system validated, and how long does validating it take us?

FDA expects you to validate the system as you use it under 11.10(a), so a vendor's own testing is the start, not the end. For Matrix Req we provide a validation package with plans, reports, test scripts and traceability matrices (Matrix Req product page, read 10 November 2026), and customers complete it in about 4 hours on average, a figure from our Matrix Req demos in April 2026.

Where is our data hosted, and will an auditor accept it?

Matrix Req runs on ISO 27001:2022 certified cloud hosting with full history on every item (Matrix Req product page, read 10 November 2026). An auditor will look for what Part 11 asks: accurate copies on request under 11.10(b), records retrievable for the retention period under 11.10(c) and a secure audit trail under 11.10(e), backed by your own validation and procedures.

Does 21 CFR Part 11 apply to a paper record sent by email?

No. Section 11.1(b) excludes paper records that are, or have been, transmitted by electronic means. A signed paper form scanned and emailed stays a paper record. If the electronic copy becomes the record you rely on for a predicate rule, it falls under Part 11.

Do we have to tell FDA we use electronic signatures?

Yes. Section 11.100(c) requires a certification, before or at the time of first use, that your electronic signatures are the legally binding equivalent of handwritten ones. It is signed by hand and, since the 2 March 2023 amendment, can be sent in electronic or paper form; the regulation points to FDA's Letters of Non-Repudiation Agreement page for where to send it.

Is 21 CFR Part 11 the same as EU GMP Annex 11?

No. Part 11 is a US regulation in the Code of Federal Regulations; Annex 11 is the EU GMP annex on computerised systems. They overlap on validation, audit trails and signatures, and the draft revision of Annex 11 asks for owned system requirements, stricter audit trails and full re-authentication at signing (Matrix Validate product page, read 10 November 2026). Our EU GMP Annex 11 page covers the EU side.

Does Part 11 apply to systems installed before 1997?

Part 11 took effect on 20 August 1997. FDA's 2003 scope and application guidance applies enforcement discretion to legacy systems operating before that date under the conditions it sets out, but the records in them must still meet the predicate rules.

Does a vendor's Part 11 claim make us compliant?

No. A vendor can build the technical controls, such as audit trails, authority checks and two component signatures. Clauses 11.10(i) to (k) and section 11.100(b) and (c) need your own training, written policies, identity checks and FDA certification, so compliance always sits with the regulated company.

Written by
Eva Kautenburger
CCO

Eva Kautenburger is Chief Customer Officer at Matrix One, where she leads Customer Success & Supp across the full portfolio of regulatory and quality management solutions for the medical device industry. A certified I. and II. Party Auditor with deep expertise in ISO 13485, EU MDR/IVDR, IEC 62304, and 21 CFR Part 820, she brings both the technical fluency and regulatory grounding that MedTech customers need to navigate complex compliance landscapes. In her role, Eva oversees a cross-functional team of Solution Consultants, Solution Engineers and Account Managers, driving onboarding, retention, support and strategic growth for customers ranging from emerging device companies to global enterprises as well as consulting intiatives to support customers in their regulatory journey.

View profile →

Request a demo and get started today.

See how Matrix One connects your requirements, risks, tests, and documentation in one platform.

Request A Demo Today

Thank you

A member of our team will be in contact within 48 hours.